EXO platform intelligent sensors represent the cutting edge in environmental monitoring technology. Manufactured with aerospace-grade titanium and wet-mateable connectors, these devices are designed to operate autonomously within a multiparameter sonde. Each unit possesses its own processing power and memory, allowing calibration to be performed remotely and remain stable during prolonged field expeditions, ensuring that data collection is as fluid as the environment being studied.
Thanks to the EXO platform, these sensors redefine field efficiency. Featuring on-board memory and processing, they allow researchers to calibrate equipment in the laboratory and distribute it across various monitoring sites without losing settings. Smart port technology protects system integrity by automatically managing power consumption, ensuring your focus remains on exploration and water quality analysis in its purest state.
| Parameter / Variant | Measurement Range | Accuracy and Resolution |
| Conductivity (C/T Sensor) | 0 to 200 mS/cm | ±0.5% of reading or 0.001 mS/cm (0-100 mS/cm) |
| Temperature (C/T Sensor) | -5 to 50 °C (23 to 122 °F) | ±0.01 °C (-5 to 35 °C); ±0.05 °C (35 to 50 °C) |
| pH (Unguarded / Guarded) | 0 to 14 pH units | ±0.1 pH within 10 °C of calibration temperature |
| ORP (Redox Potential) | -999 to 999 mV | ±20 mV in standard Redox solutions |
| Dissolved Oxygen (Optical) | 0 to 50 mg/L; 0 to 500% air | ±0.1 mg/L or 1% of reading (0 to 20 mg/L) |
| Turbidity | 0 to 4000 FNU | 0.3 FNU or ±2% of reading (0 to 999 FNU) |
| Chlorophyll (Total Algae) | 0 to 400 μg/L Chl | Linearity r2 > 0.999; Resolution 0.01 μg/L Chl |
| Phycocyanin (Freshwater PC) | 0 to 100 μg/L PC | Linearity r2 > 0.999; Resolution 0.01 μg/L PC |
| Phycoerythrin (Saltwater PE) | 0 to 280 μg/L PE | Linearity r2 > 0.999; Resolution 0.01 μg/L PE |
| Salinity | 0 to 70 ppt | ±1.0% of reading or 0.1 ppt (whichever is greater) |
| TDS | 0 to 100,000 mg/L | Variable Resolution; Cal Constant 0.30 to 1.00 |