HOBO MX800 Series multiparameter wireless data loggers work with interchangeable sensors to provide a highly configurable solution for monitoring water conductivity/salinity, temperature, depth, and dissolved oxygen. Available in both an easy-to-deploy, fully submersible model (MX801) and a direct read model (MX802) for convenient data offload that doesn’t require having to remove sensors from the water.
Specification | MX801 | MX802 |
---|---|---|
Operating Range | -40° to 50°C (-40° to 122°F), cannot be frozen in ice | -40° to 50°C (-40° to 122°F), cannot be frozen in ice |
Transmission Range | Approximately 30.5 m (100 ft) line-of-sight | Approximately 30.5 m (100 ft) line-of-sight |
Wireless Data Standard | Bluetooth 5 | |
Maximum Cable Length | N/A | 120m (400 ft) |
Logging Rate | 5 second to 18 hours | |
Water Detect | Events are logged when logger is submerged or removed from the water | N/A |
Time Accuracy | ±1 minute per month 0° to 50°C (32° to 122°F) | |
Barometric Pressure (Measurement & Calibrated Range) | N/A | 66 to 107 kPa (9.6 to 15.5 psia), -20° to 50°C (-4° to 122°F) |
Barometric Pressure (Accuracy) | N/A | Typical error: ±0.2 kPa (0.029 psi); Maximum error: ±0.5 kPa (0.073 psi) |
Barometric Pressure (Resolution) | N/A | <0.01 kPa (0.0015 psi) |
Battery | Four AA, 1.5 V lithium batteries, user replaceable | |
Battery Life | With W-CTD or W-CT sensors: 2 years, typical with 1-minute logging interval or slower. With W-DO sensor: 1 year, typical with 5-minute logging interval or slower. |
|
Memory | 730,000 measurements, divided between channels (4 Mbyte memory) | |
Materials | Outer enclosure: Polycarbonate/PBT blend with brass inserts; Interior: Polycarbonate/PBT; Gasket: Silicone foam. Cable channel: Santoprene™ TPE. |
|
Dimensions | Logger diameter: 4.14 cm (1.63 inches) Length without sensor: 20 cm (8 inches) |
|
Weight | 213g (7.51 oz) | 215g (7.58 oz) |
Environmental Rating | IP68; waterproof to 100m (with sensor(s) attached) | IP67, NEMA 6; Weatherproof (with cable or Termination Plug attached); not waterproof |
CE & UKCA Compliant | Yes |
Specification | Details MX801 AND MX802 |
---|---|
Operating Range | -40° to 50°C (-40° to 122°F), cannot be frozen in ice |
Radio Power | 1 mW (0 dBm) |
Transmission Range | Approximately 30.5 m (100 ft) line-of-sight |
Wireless Data Standard | Bluetooth 5 |
Number of Sensors That Can Be Connected | 2 |
Maximum Cable Length | 120m (400 ft) |
Logging Rate | 5 seconds to 18 hours |
Logging Modes | Fixed interval (normal or statistics), burst, or multiple intervals with up to 8 user-defined logging intervals and durations, with pause and repeat options |
Start Modes | Immediate, next interval, or date & time |
Stop Modes | Memory full, never stop (overwrite oldest data), date & time, or after a set logging period |
Water Detect | Events are logged when logger is submerged or removed from the water (MX801 only) |
Time Accuracy | ±1 minute per month 0° to 50°C (32° to 122°F) |
Battery | Four AA, 1.5 V lithium batteries, user replaceable |
Battery Life | With W-CTD or W-CT sensors: 2 years (typical with 1-minute logging interval or slower) With W-DO sensor: 1 year (typical with 5-minute logging interval or slower) Faster logging, burst mode, or excessive readouts will impact battery life |
Memory | 730,000 measurements (4 Mbyte memory), divided between channels |
Data Download Time (with Bluetooth 5+ devices) | Full memory: Approximately 3.5 minutes 120,000 measurements: 1 minute May take longer if the device is farther from the logger |
Specification | Details Barometric Pressure (MX802 only) |
---|---|
Barometric Pressure (MX802 Only) | - |
Measurement & Calibrated Range | 66 to 107 kPa (9.6 to 15.5 psia), -20° to 50°C (-4° to 122°F) |
Accuracy* | Typical error: ±0.2 kPa (0.029 psi) Maximum error: ±0.5 kPa (0.073 psi) |
Resolution | <0.01 kPa (0.0015 psi) |
Pressure Response Time** | <1 second to 90% at a stable temperature |
Contribution to Water Level Accuracy*** | Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water |
Specification | Details MX801 and MX802 Specs, continued: |
---|---|
Dimensions | - |
Logger Diameter | 4.14 cm (1.63 inches) |
Length Without Sensor | 20 cm (8 inches) |
Length with W-CTD Sensor | 36 cm (14 inches) |
Length with W-CT Sensor | 33 cm (13 inches) |
Mounting Loop | 12.7 x 22.6 mm (0.5 x 0.89 inches) |
Weight | - |
MX801 Logger Head Only | 224 g (7.90 oz) |
MX802 Logger Head Only | 226 g (7.97 oz) |
Wetted Materials | Logger: PVC housing with polyester film label, PET sensor connector end cap and mounting handle with EPDM and Buna-N O-rings, passivated 316 stainless steel hardware rated for use in saltwater; MX802 also has ePTFE vent Collar: HDPE, titanium screws Note: Logger should be mounted so that no metal parts of the logger are in contact with other metals |
Depth/Environmental Rating | - |
MX801 | IP68; waterproof to 100m (with sensor(s) attached) |
MX802 | IP67, NEMA 6; Weatherproof (with cable or Termination Plug attached); this model is not waterproof |